Prerequisites
1 - 3
Siemens AG
SPR2-230.061.01
Page 3 of 18
SIREMOBIL Iso-C
Medical Solutions
Rev. 13
12.04
CS PS 24
System Manual
Safety Information
1
Danger of possibly fatal physical injury and property damage!
Observe
- the product-specific safety information in these instructions,
- the general safety information in the instructions
TD00-000.860.01... and
- the general safety information according to ARTD Part 2.
Noncompliance can lead to death, physical injury or property
damage.
Electrical voltage!
Noncompliance can lead to death, physical injury or property
damage.
Once the system covers have been removed, voltage-conducting
parts become contactable. To prevent danger, always disconnect
the system from the line voltage before removing the covers.
To do this, pull the power plug or switch the system power supply
off and secure the power switch against accidental activation.
If an uninterruptible power supply (UPS) is installed in the
system, the power supply of the UPS must be switched free of
voltage in addition or the voltage output plug must be pulled out.
If service in connection with electrical voltage is required,
the general safety information in instructions TD00-000.860.01...
must be observed.
Electrical voltage!
Noncompliance can lead to death, physical injury or property
damage.
After service is performed on the primary circuit of the ON/OFF
assembly (e.g. replacement of the ON/OFF assembly or any of its
components) the device leakage current must be measured and
recorded (see system manual or logbook).
Electrical voltage
!
Noncompliance can lead to death, physical injury or property
damage.
After completing service work and refitting all covers, a protective
ground wire test must be performed according to
ARTD-002.731.17.. . The protective ground wire resistance must
not exceed 0.2 ohms.
WARNING
WARNING
WARNING
WARNING