SIREMOBIL Iso-C
SPR2-230.061.01
Page 42 of 48
Siemens AG
System Manual
Rev. 13
12.04
CS PS 24
Medical Solutions
6 - 42
Replacing boards / replacing components
Orbital drive toothed belt
6
•
Loosen the belt tensioner of the toothed-belt orbital drive (1/Fig. 17).
•
Disconnect the connecting cables of the orbital motor (motor incremental en-
coder).
•
Loosen the 2 fastening screws of the mounting plate (incl. motor) (1/Fig. 18).
•
Loosen the 2 fastening screws at the two lateral brackets (1/Fig. 19).
•
Carefully remove the mounting plate incl. motor. In doing so, push the toothed belt off the
pulley, ensuring that the actual value potentiometer and its pulley are not damaged or
bent.
•
Unscrew the two lateral brackets from the coupling unit (3 Allen screws each,
(2/Fig. 19).
Fig. 17
Fig. 18
Fig. 19
Fig. 20
Only applies to SIREMOBIL Iso-C 3D.
1
1
1
1
1
1
2
2
NOTE