Replacing boards / replacing components
•
Secure the camera optics again.
•
Erase the three markings on the monitor.
•
Switch fluoro on briefly and mark the new 0° position.
•
Rotate the camera to the -220° position.
•
Mark the center on the monitor.
•
Rotate the camera to the + 220° position.
•
Mark the center on the monitor.
•
Place an 8 mm washer over the three markings. All three markings must be located with-
in the 8 mm diameter of the washer. An 8 mm washer has a somewhat larger inner diam-
eter of 8.4 mm, therefore the markings must clearly lie within the inner diameter and
should not touch the edge of the washer. If necessary, repeat the adjustment until the
three markings fall within the 8 mm diameter.
•
Secure the camera optics.
•
Retighten the excenter screws of the camera optics.
Checks and adjustments
6
•
Check the camera rotation and adjust it if necessary.
•
Check the setting of the radiation iris and readjust it if necessary.
•
Check the display of the blades on the monitor and adjust if necessary.
•
Check the dose rate again and adjust it if necessary.
•
Perform the IQ quick test.