Replacing boards / replacing components
6 - 11
Siemens AG
SPR2-130.061.02
Page 11 of 28
SIREMOBIL Compact
Medical Solutions
Rev. 11
09.04
CS PS 24
System Manual
•
Remove the cover over the high voltage panel (C-arm away from you, I.I. near you).
•
Remove the high voltage panel from the C-arm and remove the anode cable.
•
Short out the residual charge to ground.
•
Remove the O-ring and the cap nut from the anode cable.
•
Withdraw the anode cable upward from the C-arm.
•
Place the high voltage panel back in the C-arm for now.
•
Disconnect the plugs of the control panel of the I.I. mini voltage supply (1/Fig. 5).
•
Disconnect the protective conductor connections and the ground screw (5/Fig. 5).
•
Remove plug D4.X1 of the VIDEOMED DC (2/Fig. 6).
•
Secure the cable rolled-up in the cable spool of the VIDEOMED DC with adhesive tape.
•
Remove plug X45 of the cassette contact (2/Fig. 6).
•
Remove jacks D4.X1 and X45 from the clips (2/Fig. 6).
•
Loosen the retainer clip above the camera and remove it (2/Fig. 5 or 2/Fig. 8).
•
Remove the control panel from the 9"/23 cm I.I.
•
Disconnect the plug from the VIDEOMED DC and the Compact optics.
•
Cut off the cable ties.
•
Loosen the camera set screw slightly and remove the camera (1/Fig. 7).
•
Disconnect plug M30.X1 (4/Fig. 7).
•
Loosen and remove the Compact optics (3/Fig. 7).
•
Remove the rubber gasket.
•
Remove the anode cable from the recess in the Compact optics.
Mounting the Compact optics
6
•
Reinsert the rubber gasket between the I.I. output and the camera optics. Reattach the
Compact optics to the new I.I.
Insert the anode cable through the opening in the optics.
The Compact optics must lie against the excenter screws.
•
Adjust the camera rotation of the Compact optics to the "0" position (6/Fig. 6).
•
To do this, connect an ohm meter to the potentiometer for camera rotation, location 2
and 3 (5/Fig. 7). The resistance must measure 5000 ohms + 40 ohms in the middle posi-
tion.
Do not loosen the eccentric screws (2/Fig. 7) on the edge of the
Compact optics. These are used to center the Compact optics to
the I.I. output. When removing the Compact optics, make sure
that no dust or dirt contaminates the I.I. output window or the
Compact optics.
When installing the Compact optics and the VIDEOMED DC, make
sure that no dust or dirt contaminates the I.I. output window or
the Compact optics.
NOTE
NOTE