
SIREMOBIL Compact
SPR2-130.061.02
Page 6 of 28
Siemens AG
System Manual
Rev. 11
09.04
CS PS 24
Medical Solutions
6 - 6
Replacing boards / replacing components
•
Check the setting of the laser targeting device and, if necessary, adjust it according to
the Adjustment Instructions for the Laser Targeting Device, RXR2-130.032.01.
•
Re-attach the SIREPHOS cover.
•
Seal the SIREPHOS cover using sealing compound. See Chapter 1 "Covers".
Replacing the I.I.
6
Checking the temperature indicator
6
•
Prior to installing the I.I., check the temperature indicator:
ê
If the inner square field of the indicator is white, the I.I. did not exceed the tem-
perature range. Remove the temperature indicator.
ê
If the indicator is discolored (inner field is black), proceed according to the
IQ document RXD-000.038.01
.
Fig. 1
Fig. 2
Fig. 3
Fig. 4
3
6
3
2
5
4
5
1
5
1
2
4
4
3
5
3
1
2
3
1
3
2
4
2
5
3
3