Important functions (selection)
7.3 Calibration and validation functions
SIPROCESS UV600
98
Operating Instructions, 10/2013, A5E03529375-02
7.3.2.3
Configuring calibration gases
Basics
The calibration gas settings are the basis of the calibrations. The setpoint (concentration)
and the sequence parameters for the calibration procedure are defined for a calibration gas
in a calibration gas setting. The calibration procedures for which a calibration gas setting can
be used are also defined.
It is possible to use a real calibration gas in several calibration gas settings. In this manner a
specific calibration gas can be configured with different settings for different calibrations.
All calibration gas settings can also be provided for validation measurements. A total of 12
different calibration gas settings can be configured.
Appropriate calibration gas settings are usually preprogrammed by the manufacturer. If your
operating procedure should require different calibration test settings, you can use the
following overview as support. We recommend that you use each calibration gas setting for
only one specific calibration or validation.
Overview of the calibration gas configuration options
N
o.
Menu title
Configuration
Described in
section
1 Name
Specify name of calibration gas setting
2 Active
Set availability of calibration gas setting
3 Pump off
Set pump mode
4 Valve
Select valve control
5 Use
Call application menu of a measured component
6 Component
Assign measured component
7 Active
Set availability for the measured component
8 Concentration
Set setpoint for a measured component
9 As zero gas
Set application as zero gas
10 As ref. gas
Set application as reference gas
Basic settings
You can use this function to make the basic settings of the calibration gas configuration.
Menu call:
Main menu → Calibrations → Settings → Calibration gases