Startup
5.4 Basic parameter assignment
SIPROCESS UV600
50
Operating Instructions, 10/2013, A5E03529375-02
5.4.5
Displaying/adjusting the setpoints of the calibration gases
This function is used to display the setpoints (concentrations) of the calibration gases and to
change them if required. Proceed as follows to call this menu:
1.
Activate user level 1 as described in Changing the access privileges (Page 48)'.
2.
Select the following menu sequence:
Main menu → BCU (control unit) → Parameterization → Test gases.
A menu is displayed with the currently set values.
Figure 5-4
Menu 'Test gases' (example)
3.
In order to change the value of a parameter, move the cursor (line displayed inverted)
using the
and
keys to the desired parameter and select this by pressing the
key.
The following menu is displayed:
Figure 5-5
Adjustment of setpoint (example: NO)
4.
Change the display parameter to the desired value as follows:
–
Press the
key until the correct value is displayed.
–
Press the
key in order to move the cursor on by one position.
5.
Save the desired setting using the
key.