Mounting and connecting
4.5 Electrical connections
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
31
4.5.1.2
Measures of the operating company
The safety of a system in which the analyzer is integrated is the responsibility of the
operating company.
External line fuse
An external line fuse must be provided in the power supply. The recommended rating for the
device is 16 A slow-blow.
Note
When switching-on, the SIPROCESS UV600 briefly requires a current larger than the rated
current (approximate values: 30 A with 230 V AC or 60 A with 115 V AC power supply).
Therefore fuses with slow-blow tripping characteristic must be used.
External power switch
WARNING
Endangering of electric safety resulting from live power supply
If the power supply to the analyzer or to the cables cannot be switched off using a
disconnector, this could result in an electrical accident.
•
Install a mains disconnection device with which the power supply to the device can be
interrupted in accordance with DIN EN 61010 in the direct vicinity of the device (see
rating plate for possible load). This mains disconnection device must be readily
accessible at all times.
•
Identify this mains disconnection device distinctly and unambiguously.
•
Note that the power supply may only be switched on again following completion of
connection/maintenance work by the associated personnel with observation of the
applicable safety regulations.
WARNING
Endangering of electric safety resulting from power cord of incorrect rating
When using a removable power cord, an electrical accident could occur if the specifications
are not correctly adhered to.
Always observe the exact specifications in the manual when using a removable power cord
(see section ).