Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
149
Selecting the electronic switching logic
In the previous menu, select the 'Output' parameter. The following menu is displayed
(example):
The parameters have the following meanings:
Direct
The digital output is electronically activated (relay is pulled-up) if the source
is also in the logical activated status (open-circuit principle).
Inverted
The digital output is electronically activated if the source is in the logical
deactivated status (closed-circuit principle).
If the source is logical activated, the switching output is electronically in the
deactivated status (relay dropped-out)
Set the desired switching logic using the
,
and
keys.
7.5.8.4
Configuring the analog inputs
The functionality of the analog inputs can only be configured using the 'SOPAS ET' PC
software /7/. There are no menu functions for this.