Important functions (selection)
7.5 Parameter assignments
SIPROCESS UV600
132
Operating Instructions, 10/2013, A5E03529375-02
7.5.3
Configuration of measured-value display
Number of decimal places
Up to 6 digits are available for displaying the measured value. If a measured value includes
decimal places, you can individually select the desired number of decimal places for each
measured component.
When configuring the measured-value display, please note the following:
●
If the measured-value display has more than 3 digits, the value is displayed more
accurately than the actual measuring accuracy. In addition, the last digits of the display
may change frequently even though the measured value - with consideration of the
possible accuracy - is more or less constant (measured-value noise). You can influence
this effect by means of damping (see section ).
●
If you limit the number of decimal places such that the display only has 2 or 3 digits, it
may be the case that changes in measured value are not recognized in time.
Menu call:
Main menu → Parameters → Measured
-
value display → [Measured component] → Format
The following menu is displayed (example):
Use the
,
and
keys to set the desired state.