Important functions (selection)
7.4 Maintenance functions
SIPROCESS UV600
Operating Instructions, 10/2013, A5E03529375-02
121
7.4.2
Maintenance status on/off
This menu activates the status flag "C (Check)". If a digital output is connected accordingly,
a signal can then be sent to an external system indicating that the SIPROCESS UV600 is
not in measuring mode, for example because maintenance work is currently being carried
out.
Note
We recommend that you set this status flag not only when carrying out direct maintenance
work on the analyzer but also when working on the upstream gas preparation since changes
to the system components can also influence the measurements.
Menu call:
Main menu → Maintenance → Maintenance signal
The following menu is displayed:
The parameters have the following meanings:
On: Maintenance signal activated
Off: Maintenance signal switched off
Select the desired setting using the
and
keys, and confirm the selected setting
with
.
Note
The maintenance flag can only be deactivated if it was previously set in this menu. If there
are other reasons for the "Maintenance" status, the status remains set until the reasons have
been eliminated.