
Safety notes
2.3 Proper use
SIPROCESS UV600
12
Operating Instructions, 10/2013, A5E03529375-02
2.3
Proper use
The SIPROCESS UV600 gas analyzer measures the concentration of one gas or several
gases in a gas mixture.
The following conditions apply when selecting the location of use of the analyzer and the
condition of the supplied sample gas:
Location of use
●
The SIPROCESS UV600 may only be used in closed rooms
●
The SIPROCESS UV600 must not be used in potentially explosive atmospheres
Sample gas
The SIPROCESS UV600 is not suitable for measuring flammable and/or explosive gases
and gas mixtures!
In addition, the sample gas must be free from:
●
Substances which could attack the materials of the gas path in the analyzer
●
Particles which could deposit in the gas path in the analyzer
●
Gas components which could condense in the analyzer during normal operation
Additional information is available in the Auto-Hotspot . In case of doubt, we recommend that
you discuss your application with our technical department.
It is also possible that components may be present in the sample gas which influence the
measurement (cross-sensitivity). In such a case, the displayed value deviates from the
actual measured value. In order to minimize or eliminate cross-sensitivities, we recommend
that you discuss your application with our technical department