4.3
Gas supply/discharge lines
WARNING
Suitability of gas supply/discharge lines
Unsuitable cables may result in personal injury and/or property damage. The following applies
if you use the device in hazardous areas:
● Make sure that all gas lines satisfy the requirements for operation in hazardous areas.
Note
Incorrect measured values due to burning off of sample gas
When burning off sample gas, pressure variations in the exhaust gas line may falsify the results.
We therefore recommend the following measures:
● Install a pneumatic lowpass: install a damping vessel with downstream restrictor (volume
> 1 l) between the device and the exhaust line.
● Make sure that a technical expert for hazardous areas examines and approves the
measures. Only commence operation of the device following the approval.
Application planning
4.3 Gas supply/discharge lines
Field device
38
Operating Instructions, 05/2015, A5E35640463-02
Summary of Contents for SIPROCESS GA700
Page 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Page 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Page 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Page 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Page 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...