
Operator control and monitoring
9.2 Working with the user interface
SIMATIC MV420 / SIMATIC MV440
Operating Instructions, 04/2013, A5E02371045-06
219
Settings and programs
With the "Save/restore" button, you open a Web page on which you can save the saved
programs and parameter records or reload them to the reader.
Firmware
Note
Control via automation system: DISA bit must not be set
If you control the reader from an automation system, make sure that the DISA bit is not set
during the firmware update. This prevents an automation system from switching the reader
to RUN.
"Incremental update..." button
The "Incremental update..." button opens a Web page where you can update the firmware.
To do this, you require a vsz file.
An incremental update is only possible to a version in which the version has only changed in
the third digit compared with the previous version.
Summary of Contents for SIMATIC MV420
Page 2: ......
Page 418: ...Glossary SIMATIC MV420 SIMATIC MV440 416 Operating Instructions 04 2013 A5E02371045 06 ...
Page 423: ......