Operation
8.4 Menu-based operation
Set CEM CERT
Operating Instructions, 06/2021, A5E34491309-12
59
8.4.2.3
Language/Time of day
The language and the time of day can be set in this menu.
The following operating languages are currently supported:
•
German
•
English
•
French
When the "Set time" parameter is selected, the time setting is updated on the following
components:
•
SIMATIC controller
•
SIMATIC HMI
•
Following devices of Series 6
–
ULTRAMAT 6
–
OXYMAT 6
–
FIDAMAT 6
•
SIPROCESS GA700 devices
Note
The following devices are not updated by this menu:
•
ULTRAMAT 23 This device has no internal clock.
•
LDS 6 and UV600: Here, the time of day
cannot
be set via the HMI.
The time of day must be set using the control panel of the corresponding device (see
manual of the respective device for this).