Installation
5.3 Installing the individual components
Set CEM CERT
Operating Instructions, 06/2021, A5E34491309-12
27
To reach the mounting brackets, you may have to open the floor covering (arrow).
5.3
Installing the individual components
5.3.1
Installation of the sampling probe
The gas sampling probes are designed for flange mounting.
Their installation location and position result from the application-relevant requirements.
If possible, the probe installation fitting should be at a slight incline from the center of
the channel. The installation location should be weather-protected (heat, cold,
vibrations).
Observe the information and directions in the relevant operating instructions when
installing the sampling probes.