Set CEM CERT
Operating Instructions, 06/2021, A5E34491309-12
25
Installation
5
5.1
Preparing the installation location
The operator is responsible for preparing the installation location.
We recommend that the analysis cabinet be installed in a dust-protected control room
and that the following ambient conditions be observed:
•
Level installation surface with adequate load bearing capacity (> 500 kg/m
2
)
•
Protection against direct thermal radiation
•
Protection against heavy dust load
•
Protection against corrosive atmosphere
•
Protection against vibrations, as these can influence the measurement
Therefore, in case of doubt, you should provide vibration damping on site
•
Installation of the measuring system as close as possible to the point of
measurement, because short sample gas lines are essential for achieving short T
90
times
Note
Follow the installation instructions for installing heated sample gas lines (e.g. correct
bending radius, routed with a downward slope if possible, no routing of heated sample
gas lines side-by-side).