Description
3.2 Design
Set CEM CERT
18
Operating Instructions, 06/2021, A5E34491309-12
3.2
Design
Tested component design
The overall modular measuring system consists of the following tested individual
components:
•
Analyzers for measurement of multiple components. The Set CEM CERT modular
measuring system can consist of up to three analyzers in combination with a system
cabinet. The performance-tested analyzer(s) are selected separately from the system
using the specific article number(s). The system is certified for use of the following
analyzers:
Analyzer
Article No. Design
ULTRAMAT 23*
7MB2358
3 NDIR components on 2 optical benches or
2 NDIR and 2 UV components on 2 optical benches
ULTRAMAT 23*
7MB2357
2 NDIR components on 2 optical benches or
1 NDIR and 2 UV components on 2 optical benches
ULTRAMAT 23*
7MB2355
1 NDIR component on 1 optical bench or
1 NDIR component or 2 UV components on 1 optical
bench
SIPROCESS UV600
7MB2621
3 UV components on 1 optical bench
ULTRAMAT 6
7MB2121
7MB2011
1 NDIR component on 1 optical bench
ULTRAMAT 6,
2-channel 19" rack unit
7MB2123
7MB2124
2 NDIR components on 2 optical benches
OXYMAT 6
7MB2021
1 paramagnetic O
2
sample chamber
ULTRAMAT/
OXYMAT 6
7MB2023
7MB2024
1 NDIR component on 1 optical bench and
1 paramagnetic O
2
sample chamber
LDS 6**
7MB6121
TDLAS diode laser, consisting of a central unit and CD 6
in-situ cross-duct sensors
* Each measuring module of the ULTRAMAT 23 analyzer type can be additionally equipped with
different O
2
sensors. The configuration with an O
2
sensor is specified by adding the following
identifiers:
-Z-T13/T14: Paramagnetic O
2
sensor
-Z-T23/24/T25: Electrochemical O
2
sensor
-Z-T33/T34/T35: No O
2
sensor
** The overall system consists of the following components:
- Central unit 7MB6121
- Sensor unit 7MB6122
- Hybrid cable (connects central unit to CD 6 sensor sender)
- Sensor connecting cable (connects CD 6 sensor sender to CD 6 sensor receiver)
•
Sampling probe:
–
M&C, type: SP2000
–
Bühler, type: GAS222
•
Heated sample gas line: max. length 50 m, hold temperature 180 ... 200
°C, 100
W/m
Core: PTFE DN 4/6
Summary of Contents for Set CEM CERT
Page 1: ......
Page 2: ......
Page 8: ...Table of contents Set CEM CERT 6 Operating Instructions 06 2021 A5E34491309 12 ...
Page 44: ...Commissioning 7 2 Commissioning Set CEM CERT 42 Operating Instructions 06 2021 A5E34491309 12 ...
Page 112: ...Technical specifications Set CEM CERT 110 Operating Instructions 06 2021 A5E34491309 12 ...
Page 122: ...Appendix A 2 Returns Set CEM CERT 120 Operating Instructions 06 2021 A5E34491309 12 ...