Set CEM CERT
Operating Instructions, 06/2021, A5E34491309-12
105
Alarm, fault and system messages
10
10.1
Error messages
A group fault message is output on the operator display if one or more of the following
conditions is not met:
•
All fuses are OK
•
All gas preparation elements are OK
•
The maximum permissible cabinet temperature has not been reached or exceeded
•
The minimum temperature of the sampling probe has been reached
•
The minimum temperature of the sample gas line has been reached
•
All air conditioning units (options) are OK
•
Analyzer 1 is OK
•
Analyzer 2 (option) is OK
•
Analyzer 3 (option) is OK
•
Flow rate is in the permissible range (72 ... 120 l/h) for each analyzer
The group fault message remains active until all individual fault messages are
eliminated. The fault messages must be acknowledged on the SIMATIC HMI display
after correction of the fault.
The group fault message is also output when the system is switched on. It is normally
reset automatically after the warm-up period ends.
The following errors are reported:
Message
Meaning
Fault -EC1 sample gas cooler T > max
Fault sample gas cooler (see corresponding
manual)
Fault -EC2 air conditioning unit
Fault air conditioning unit (see corresponding
manual)
Fault -BC1 converter
Fault NOx converter (see corresponding
manual)
Fault -BS3 control cabinet temperature T > max
Temperature in control cabinet too high
Fault -EH1 heated sample gas line T < min
Temperature of sample gas line too low
Fault probe -WS1 temperature T < min
Temperature of probe too low
Fault -F... miniature circuit breaker tripped
One or more fuses tripped
Analyzer -BAx: Fault
Fault on analyzer x
Analyzer -BAx: CHy fault
Fault in measurement channel y on analyzer x
Analyzer -BAx: Function check
Calibration process active or operator control
of analyzer BAx
Analyzer -BAx: CHy function check
Calibration process active or operator control
of measurement channel y of analyzer x
Summary of Contents for Set CEM CERT
Page 1: ......
Page 2: ......
Page 8: ...Table of contents Set CEM CERT 6 Operating Instructions 06 2021 A5E34491309 12 ...
Page 44: ...Commissioning 7 2 Commissioning Set CEM CERT 42 Operating Instructions 06 2021 A5E34491309 12 ...
Page 112: ...Technical specifications Set CEM CERT 110 Operating Instructions 06 2021 A5E34491309 12 ...
Page 122: ...Appendix A 2 Returns Set CEM CERT 120 Operating Instructions 06 2021 A5E34491309 12 ...