3
96
Siemens-Elema AB
E382 E380E 061 01 03 01
Disassembling and assembling
Servo Ventilator 300/300A
Assembling the power
section
The procedure for assembling the power
section is the reverse of the disassembly
procedure previously described.
The following points must be noted during
assembling:
Mains power inlet
• Connect the cables to the mains power
inlet according to figure.
1 2 3 4
RD
WH
BK
GN
300-A40X
Power supply board PC 1618
• Connect cables to screw terminal
according to figure.
RD
BU
BK
BK
VT
BN
YE
WH
300-A39X
A
C
E
B
D
F
G
H