1-56
Rapidlab 1200 Operator’s Guide: System Overview and Intended Use
02087462 Rev. V
Patient Temperature Correction
The Rapidlab 1200 systems measurements and determinations are based on a standard
temperature of 37.0
°
C. During sample analysis, you can enter the actual patient
temperature value, which enables the system to provide temperature-corrected results for
pH,
p
CO
2
, and
p
O
2
.
The system determines temperature-corrected results using the following correction
factors:
19
pH correction =
Δ
pH /
Δ
T = –0.0147 + 0.0065(7.4–pH)
0.019
T
CO
log
correction
CO
2
2
=
Δ
Δ
=
p
p
2.30
O
10
9.72
0.071
O
10
5.49
T
CO
log
correction
CO
3.88
2
9
–
3.88
2
–11
2
2
+
×
×
+
×
×
=
Δ
Δ
=
p
p
p
p
Hemoglobin Oxygen Saturation
Hemoglobin oxygen saturation (
s
O
2
) is a ratio of the amount of hemoglobin bound to
oxygen to the total amount of hemoglobin able to bind oxygen.
19
Hemoglobin oxygen
saturation, with oxygen content and oxygen capacity, is a useful parameter for determining
the amount of oxygen in the blood that is actually available to the tissues and for
determining the effectiveness of oxygen therapy.
Hemoglobin oxygen saturation, expressed as a percent, is determined using the following
equation:
s
O
2
= (100 x
F
O
2
Hb) / (
F
O
2
Hb +
F
HHb)
Oxygen Content
Oxygen content is the concentration of the total oxygen carried by the blood, including
oxygen bound to hemoglobin as well as oxygen dissolved in plasma and in the fluid within
red cells.
Clinically, dissolved oxygen is unimportant for most situations. However, at very low
levels of hemoglobin or in patients receiving hyperbaric oxygen therapy, dissolved oxygen
may be a very significant contributor to oxygen content and thus to oxygen transport.
Oxygen content is determined, using CLSI recommendations, from the following
relationship:
c
tO
2
= (
F
O
2
Hb ×1.39 ×
c
tHb + 0.00314 ×
p
O
2
)
where
c
tHb is expressed in g/dL and
F
O
2
Hb is decimal.
19
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...