Rapidlab 1200 Operator’s Guide: System Overview and Intended Use
02087462 Rev. V
CO-ox Sample Path
The system splits the sample in the preheater and moves part of the sample through the
CO-ox module. The CO-ox sample moves through the CO-ox sample tubing to the sample
chamber for analysis and exits as waste through the waste tubing. The system analyzes the
sample in the CO-ox sample chamber. After analysis, the system moves the waste to the
reagent manifold, passing through the reagent cartridge, and on to the waste bottle.
Figure 1-27 Rapidlab 1200 Sample Path–To CO-ox Module to Waste
1
Preheater
2
Sample chamber
3
Tubing to the reagent manifold