E-66
Rapidlab 1200 Operator’s Guide: System Specifications
02087462 Rev.V
Recovery and Precision of Hemoglobin Fractions in Human
Whole Blood
For testing capillary, tHb, and syringe modes, blood was collected in heparinized vacuum
tubes. It was tonometered at 37°C or adjusted chemically. Three target levels of tHb were
prepared by separating the red cells from the plasma and recombining appropriately. For
each run, 3 replicates of each level were run on a minimum of 3 Rapidlab 1265 systems.
The mean bias was determined by taking the difference between the Rapidlab 1265 system
mean and the reference methods mean. The mean bias and the pooled within run standard
deviation were determined for each blood preparation and sampling mode.
The following tables summarize the results of the Rapidlab 1265 CO-ox module recovery
and precision testing.
Table E-63 Rapidlab 1265 CO-ox Module Recovery and Precision with Total
Hemoglobin of ~6 g/dL and Oxyhemoglobin > 75%
Analyte
Mode
*
* tHb mode performance is equivalent to sample device used
n
Mean
Reference
Mean Bias
WRSD
†
† WRSD: within-run standard deviation
tHb
Capillary
61
6.20
0.29
0.08
tHb
tHb
55
6.19
0.29
0.15
tHb
Syringe
58
6.16
0.23
0.05
F
O
2
Hb
Capillary
63
86.67
-1.65
0.21
F
O
2
Hb
tHb
56
87.21
-1.27
0.42
F
O
2
Hb
Syringe
60
86.46
-1.52
0.34
F
COHb
Capillary
63
5.00
0.67
0.21
F
COHb
tHb
57
4.82
0.86
0.41
F
COHb
Syringe
60
4.80
0.87
0.27
F
MetHb
Capillary
63
4.80
0.22
0.08
F
MetHb
tHb
57
5.48
-0.16
0.09
F
MetHb
Syringe
61
4.89
-0.06
0.10
F
HHb
tHb
55
2.68
-0.05
0.32
F
HHb
Capillary
63
3.54
0.29
0.14
F
HHb
Syringe
61
3.77
0.03
0.21
Table E-64 Rapidlab 1265 CO-ox Module Recovery and Precision with Total
Hemoglobin of ~14 g/dL and Oxyhemoglobin > 75%
Analyte
Mode
*
n
Mean
Reference
Mean Bias
WRSD
†
tHb
Capillary
60
14.14
0.33
0.12
tHb
tHb
62
14.17
0.18
0.14
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...