Rapidlab 1200 Operator’s Guide: System Specifications
E-49
02087462 Rev.V
Rapidlab 1265 System Performance Characteristics
All performance data presented in this section was generated using the
Rapidlab 1265 system. The system performed calibrations using the default
settings recommended by Siemens for optimum performance. All reported values
were corrected to 760 mmHg. All data was collected at room temperature (about
23°C). You should determine your own performance characteristics in your
laboratory with your Rapidlab 1265 system.
Precision on Controls
Quality control materials, including manual (QC) and automatic (AutomaticQC)
were analyzed on the Rapidlab 1265 system. Manual QC is performed by the
operator while AutomaticQC is performed automatically by the system based on
the QC frequency defined in the system software. The results are presented here.
Precision on manual aqueous quality control materials was estimated using a
minimum of 3 Rapidlab 1265 systems. Controls were measured over a minimum
of 10 days.
Precision on the aqueous quality control materials in the AutomaticQC module
was estimated using a minimum of 3 Rapidlab 1265 systems over a minimum of
10 days. AutomaticQC is automatically scheduled to run (single sample)
independently 3 times per day. The following tables summarize the results of the
Rapidlab 1265 system precision for manual QC and AutomaticQC.
pH, Glu,
Lac
Capillary
120
0.842
0.423
0.527
0.990
0.8
16.6
pH, Glu,
Lac
Syringe
230
0.857
0.326
0.545
0.990
0.7
17.8
Table E-52 Rapidlab 1265 Manual QC Precision Results
Parameter
Level
n
Mean
WRSD
*
Total SD
†
Ca
++
1
145
1.67
0.005
0.017
Ca
++
2
146
1.31
0.005
0.013
Ca
++
3
146
0.84
0.007
0.010
Cl
-
1
147
78
0.38
0.49
Cl
-
2
148
98
0.26
0.48
Cl
-
3
148
118
0.44
0.63
Glu
1
147
197
1.27
2.02
Table E-51 Rapidlab 1260 Human Whole Blood Method Comparison—Lactate
Mode
N
Slope
Intercept RMSE
r
2
Min
Max
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...