8-44
Rapidlab 1200 Operator’s Guide: System Configuration
02087462 Rev. V
Setting up Correlation Coefficients
Use this procedure to enter slope and offset (intercept) values to correlate the patient
sample results from the Rapidlab 1200 system to those obtained using another
measurement system. The following table lists the valid ranges for slope and offset:
1. Select
Status > Setup > Secured Options
.
2. Select the down arrow button and then select
Correlation Coefficients
.
3. Enter the slope and offset values for each parameter that you want to correlate to
another measurement system.
4. Select
Save
.
Setting up Maintenance Functions
The following functions are available on the Secured Options > Maintenance screens:
•
Setup Schedule
•
Export Schedule
•
Import Schedule
Setting up a Maintenance Schedule
At the Maintenance–Schedule screen, you can perform the following tasks:
•
view maintenance task details
•
add up to ten user-defined maintenance tasks
•
edit predefined and user-defined maintenance tasks
•
delete user-defined maintenance tasks
•
enable or disable some predefined and user-defined maintenance tasks
CAUTION:
Do not change the correlation coefficients until you collect the
appropriate data. Changing correlation coefficients affects the reported results for
patient samples. Before changing correlation coefficients, determine the slope and
offset values, as described in
Performing a Correlation Study
Parameter
Default
Slope
Slope
Range
Default
Offset
Offset
Range
pH
1.0
0.8–1.2
0.0
+/- 9.9
p
CO
2
,
p
O
2
, Na
+
, K
+
, Ca
++
, Cl¯, Glu,
Lac, tHb
1.0
0.8–1.2
0.0
+/- 99
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...