8-6
Rapidlab 1200 Operator’s Guide: System Configuration
02087462 Rev. V
Setting up AutomaticQC Ranges
Use the AutomaticQC ranges screen to view lot information and target ranges for controls
used for AutomaticQC analysis.
When you replace the AutomaticQC cartridge, the system replaces the current target
ranges with its own target ranges. If you want to retain the ranges that you defined,
perform the following steps. You can also use this procedure to change target ranges back
to the recommended default ranges.
1. At the Setup screen, select
QC > AutomaticQC Ranges
.
If your system is connected to a data management system or an LIS and you select an
option that changes target ranges, be sure to change target ranges at the data
management system or LIS.
2. Keep or change the target ranges.
3. Select
Save
.
Setting up High G/L QC Options
Use this screen to configure demographic data entry fields for High G/L samples. During
analysis, the system can require the operator to enter selected demographic information
for High G/L samples.
NOTE:
If both Glucose and Lactate are turned off in Setup, the Analysis screen does not
display the High G/L option.
1. At the Setup screen, select
QC > High G/L
.
2. Select either or both of the demographics you want entered when performing
High G/L samples.
•
If you want the operator to enter their identification when performing a High G/L
test, select
Operator ID
.
•
If you want the operator to enter the lot number of the control when performing a
High G/L Test, select
Lot Number
.
3. Select the Required button to make the selected demographic required.
When you select Required next to either or both of the test demographic buttons, the
operator must enter the corresponding demographic before leaving the screen.
4. Select
Save
.
If you want to...
then...
keep the target ranges you defined
for AutomaticQC,
select
Retain Ranges
.
change the target ranges back to the
recommended default ranges,
a.
Deselect
Retain Ranges
to turn the option
off.
b. Select
Reset Defaults
.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...