6-74
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
D78 No Reagent Detected at CO-ox: 4
Viewing System Messages
The system displays messages in the events log at the Status screen or in the Recall menu.
For example, after you replace an expired wash cartridge, the message about the cartridge
no longer displays at the Status screen. It does remain in the events log and you can access
it from the Recall menu. For quick access to information about system messages, select
Help
.
The following table lists the messages in alphabetical order.
Problem
Corrective Action
Fluid detector 3
(FD3), the CO-ox
sample chamber, or
both do not detect
RCx reagent during
a calibration.
•
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
•
Remove obstructions.
•
Repeat the RCx test to check flow through the sample path.
•
Perform a leak test.
Refer to
‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
•
If no other D codes occur, replace the CO-ox sample
chamber.
•
If the problem persists, call for technical assistance.
Message
Probable Cause
Corrective Action
AQC Cartridge
Expired
The AutomaticQC cartridge has
exceeded its operating life or is
depleted.
Replace the cartridge. Refer to
Replacing the AutomaticQC
Cartridge
To continue analyzing patient
samples if you are not able to
replace the cartridge, turn off the
AutomaticQC option in Setup.
For information about setting up
the AutomaticQC options, refer
to
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...