Rapidlab 1200 Operator’s Guide: Troubleshooting
6-61
02087462 Rev. V
D23 Reagent Flow Error: 4, 12
D23 Reagent Flow Error: 5, 13
Problem
Corrective Action
Slope Cal reagent
was not detected
during calibration.
4 FD1
12 FD2
1. Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
2. Perform a leak test.
‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3. If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
4. If the problem persists, call for technical assistance.
Problem
Corrective Action
RCx reagent was
not detected during
calibration.
5 FD1
13 FD2
1. Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
2. Perform a leak test.
‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3. If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
4. If the problem persists, call for technical assistance.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...