6-56
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
D6 Codes
D6 Excessive Noise:
p
O
2
p
CO
2
Problem
Corrective Action
The sensor signal
has excessive noise.
The patient sample
result for the
affected sensor is
reported as question
result (-----?).
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
•
Ensure that you did not open or apply pressure to the
measurement module door during analysis.
If you opened or applied pressure to the door during
analysis, repeat the sample analysis.
•
Ensure that you did not apply pressure to the top cover
during analysis.
If you applied pressure to the cover during analysis, repeat
the sample analysis.
•
Perform a successful calibration (1-point or 2-point).
•
If a D6 for
p
CO2, with a drift greater than 4.0 mmHg, occurs
twice in a 4-hour period and no drift occurs for
p
O2,
perform QC analysis.
If the
p
CO2 result is out of range, replace the sensor.
•
If a D6 occurs during sample analysis, repeat the sample.
•
Deproteinize the sample path.
Refer to
Deproteinizing the Sample Path
•
If the problem persists, replace the sensor.
•
If the problem persists, call for technical assistance.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...