6-36
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
Using the Fluidic Functions Screen
The Diagnostics > Fluidic Functions screen displays buttons for performing the following
actions:
•
perform Reagent Flow test for Wash, 200 Cal, Slope Cal, RCx, or Wash Cycle
•
prime the AQC Cartridge
•
perform the Leak Test
The Fluidic Function screen also displays a graphic of the Rapidlab system fluidics to
enable you to localize the site of any leak. The graphic is only displayed when the system
detects a leak after you perform the leak test.
Performing the Reagent Flow Tests
The system highlights one or more of the Fluidic Function buttons to indicate that you
should perform that test to clear a current error. The system removes the highlighting
when you perform the test successfully.
Use this procedure to test the flow of reagents from the reagent and wash cartridges.
1. Select
Status > Diagnostics
.
2. Enter your password, if necessary.
3. Select
Fluidic Functions
.
4. Select the test to perform.
The system moves the reagent through the sample path.
5. Look for obstructions or uneven flow that may indicate a problem.
6. To test another reagent, repeat this procedure from step 4.
7. Select
Return
.
8. To return to the Status screen, select
Return
.
The system performs a wash.
Test
Description
Wash (wash cartridge)
performs the wash section of a calibration
200 cal (reagent cartridge)
performs the 200 cal section of a calibration
Slope cal (reagent cartridge)
performs the Slope cal section of a calibration
RCx (wash cartridge)
performs the RCx section of a calibration
Wash Cycle (wash cartridge)
perform a wash sequence
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...