Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
b. Push the clot-removal line through the fluid connector on the reagent manifold
until the clot-removal line appears again.
Figure 6-14 Rapidlab 1200 Systems Sample Path
Reinstalling the Sensors
1. Dry each sensor, ensure that the O-rings are in place, and then reinstall each sensor.
2. Connect the pinch valve tubing to the latch and the reagent manifold.
3. Press the tab on the spring-loaded latch to release the latch.
4. Close the measurement module door.
Completing the Inspection Procedure
1. Reinstall the cartridges and close the reagent and wash doors.
2. Perform the RCx reagent flow test:
a. Select
Return
.
b. Select
Fluidic Functions > RCx
.
c. Look at the sample path for obstructions and flow problems.
3. Select
Wash Cycle
.
4. Select
Return
.
5. Select
Analyze > Yes
.
6. Verify sensor performance by analyzing a minimum of 2 levels of quality control
material.
1
Fluid connector
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...