6-28
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
d. If the obstruction is not removed, use a syringe filled with water and inject the
water through the sample path.
5. For Rapidlab 1260 and 1265 systems, inspect the Na
+
, Ca
++
, Cl¯, K
+
, Glu, and Lac
sensors for obstructions:
a. Remove the sensors one at a time and inspect each one for obstructions.
b. Fill a syringe with water, place the syringe against the sample path, and inject
water through the sample path.
c. If the obstruction is not removed from the Na
+
sensor, push the clot-removal line
through the sample path to remove obstructions.
6. Remove obstructions from the path through the latch:
a. Grasp the tab on the reference sensor and pull it up and out of the measurement
module.
b. Disconnect the pinch valve tubing from the spring-loaded latch.
CAUTION:
Do not push the clot-removal line through the Ca
++
, Cl¯, K
+
, Glu, and
Lac sensors. Pushing the clot-removal line through these sensors could cause
damage to the sensors.
CAUTION:
Do not exert excessive pressure on the glucose or lactate sensor
membrane.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...