Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
b. Push a clot-removal line through the preheater from right to left, opposite to the
direction of sample flow, until you see the clot-removal line at the sample
connector.
Figure 6-10 Rapidlab 1200 Systems Sample Path
4. Inspect the
p
O
2
and
p
CO
2
(gas), sample ground/temperature, pH, and the reference
sensors for obstructions:
a. Remove the gas sensors one at a time and inspect each one for obstructions.
1
Preheater
2
Sample connector
CAUTION:
Do not push the clot-removal line through the gas sensors. To avoid
damaging the gas sensors, do not exert excessive pressure to the gas sensor
membranes.