Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
Troubleshooting Patient Results
Use this section to identify the symbols that are displayed with results or instead of results
for a parameter.
This symbol is displayed next to results on the screen and printed reports, and the system
displays the Interfering Substance message in the event log and on printed reports.
Table 6-2 Symbols -----
↑
or -----
↓
Probable Cause
Corrective Action
The result for the parameter is
above (-----
↑
) or below (-----
↓
)
the reporting range for the
parameter.
NOTE:
The system displays
these symbols in place of a result.
1. Analyze the sample again, ensuring that the
sample has no bubbles.
Checking for bubbles is especially important
for samples introduced from capillary tubes.
2. Analyze QC samples and verify that the QC
results are within the target ranges.
Table 6-3 Symbol -----?
Probable Cause
Corrective Action
The system detected an atypical
response when measuring the
parameter.
Analyze the sample again to verify the result.
The sensor does not reach a
stable reading within the
predefined time limit.
For information about D5 No Endpoint, refer to
The sensor signal has excess
noise.
For information about D6 Excessive Noise, refer to
Table 6-4 Symbol #
Probable Cause
Corrective Action
The sample contains substances
that may interfere with glucose or
lactate measurement.
Check the sample source and result, and ascertain
whether the presence of an interfering substance is
likely.
For more information about interfering substances,
refer to
Repeated, unexpected occurrences of this message
may indicate sensor failure. Replace the sensor.
Table 6-5 Missing Results
Probable Cause
Corrective Action
The parameter was turned off at
the Analysis screen.
Analyze the sample again, ensuring that the
parameter is turned on at the Analysis screen.
The parameter was turned off in
Setup.
Ensure that the parameter is turned on in System
Setup. Refer to
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...