5-56
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Flushing the Tubing for the Measurement Module Waste
1. Disconnect the tubing on the reagent manifold coming from the pressure sensor
bubbler.
2. Connect the syringe tubing to the connector on the reagent manifold.
Figure 5-30 Reagent Manifold Connector
3. Flush the tubing with a minimum of 3 mL of water.
4. Flush the tubing with a minimum of 12 mL air.
5. Collect the water in an absorbent swab or equivalent.
6. Reconnect the tubing to the reagent manifold.
1
Connector
2
Pressure sensor bubbler
3
Water collection area