Rapidlab 1200 Operator’s Guide: Maintenance
5-51
02087462 Rev. V
Cleaning and Drying the Tubing
Ensure that you follow these instructions using at least the minimum volumes of
fluid and air described in these procedures.
Flushing the Tubing for the RCx Reagent
1. Connect a piece of tubing to a syringe.
2. Connect the syringe tubing to the connector for the RCx reagent.
Figure 5-25 Reagent Manifold
3. Flush with a minimum of 3 mL water.
4. Flush with a minimum of 12 mL air.
5. Collect the water in an absorbent swab or equivalent.
1
Water collection area
2
RCx reagent Connector