Rapidlab 1200 Operator’s Guide: Maintenance
5-39
02087462 Rev. V
Maintaining the Internal Reference Electrode
1. Remove the internal reference electrode using the hex tool.
Figure 5-20 Reference Sensor Casette
2. If you are replacing the internal reference electrode, dispose of the electrode
according to your institution’s protocol.
3. If you are replacing the internal reference electrode, rinse the electrode
compartment with KCl fill solution:
a. Empty the fill solution from the sensor.
b. Rinse the internal electrode compartment with 3 drops of KCl fill
solution.
c. Empty those drops of fill solution from the sensor.
4. Fill the internal electrode compartment with KCl solution.
Refer to
Filling the Reference Sensor Cassette
5. Reinstall the sensor.
Refer to
6. Verify sensor performance.
Refer to
1
Internal electrode
2
Cassette
3
Internal electrode compartment
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...