Rapidlab 1200 Operator’s Guide: Maintenance
5-29
02087462 Rev. V
Maintaining the Sensors
Performing regular maintenance on the sensors is important to the life of the
sensors. This section contains the following information:
•
Replacing the sensors
•
Performing reference sensor maintenance
•
Refilling the sensors
Replacing the Sensors
You need the following materials for this task:
•
Appropriate sensor
•
Appropriate sensor fill solution
•
Reference sensor replacement kit
•
Hex tool
•
Lint-free tissue
Preparing the Sensors
•
Equilibrate new Glucose and Lactate biosensors at room temperature
(18
°
to 25
°
C) for at least 1 hour before installing.
•
Keep the biosensors in their foil packages while they are equilibrating.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...