7. Perform the cleaning cycle:
a. Select
Yes
.
b. Draw the 10% bleach solution into a syringe and insert the syringe into the sample
port.
c. Select
Continue
.
d. When prompted, remove the syringe and select
Continue
.
Do not replace the test/blank sensors at this time.
8. Select
Continue
.
9. Select
Wash Cycle
.
Reinstalling the Reference Sensor and Biosensors
1. Lift the front cover and then lift the measurement module door.
2. Push the spring-loaded latch to the right.
3. Grasp the tab on the test/blank reference sensor and pull the sensor up and out.
4. Align the top of the reference sensor with the sensor contact.
5. Snap the sensor into place.
6. If you have a Rapidlab 1260 or 1265 system, replace the test/blank sensors with the
Glucose and Lactate sensors.
Completing the Sample Path Maintenance
1. Use the sensor names on the contact assembly to verify that the sensors are installed in
the correct order.
2. Press the tab on the spring-loaded latch to release the latch.
3. Close the measurement module door and then close the front cover.
4. Wait 30 minutes for the sensors to warm up.
5. Select
Return > Analyze > Yes
.
6. Analyze a minimum of 2 levels of quality control material.
7. Dispose of the bleach solution according to your institution’s protocol.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...