Rapidlab 1200 Operator’s Guide: Maintenance
5-17
02087462 Rev. V
Replacing the Reagent Manifold
You need a reagent manifold for this task.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
1. Select
Status > Diagnostics > Cartridges
.
2. Select
Eject Both Cartridges
, remove the cartridges, and set them aside.
3. Remove the reagent manifold:
a. Lift the front cover.
Refer to
Preparing for Maintenance Procedures
b. Disconnect the waste tubing from the reagent manifold.
c. For Rapidlab 1245 and 1265 systems, disconnect the reagent manifold
tubing from the CO-ox pump tubing at the fitting.
Do not disconnect the reagent manifold tubing from the reagent manifold.
Figure 5-7 Sample Chamber and Reagent Manifold
1
Fitting
2
Reagent manifold tubing (Rapidlab 1245 and 1265 systems)
3
Waste tubing
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...