Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Replacing the Measurement Module Tubing
You need a tubing kit for this task.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
1. Select
Status > Diagnostics > Cartridges
.
2. Select
Eject R Cartridge
.
3. Lift the door on the measurement module.
Refer to
Preparing for Maintenance Procedures
4. Replace the pinch valve tubing.
Perform steps 4 and 5 of
Replacing the Pinch Valve Tubing
5. Replace the waste tubing between the pressure sensor bubbler and the reagent
manifold:
a. Disconnect the waste tubing from the pressure sensor bubbler and the reagent
manifold.
Figure 5-3 Measurement Module–Rapidlab 1260 and 1265 systems
1
Pressure sensor bubbler
2
Waste tubing
3
Reagent manifold
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...