Rapidlab 1200 Operator’s Guide: Operating the System
2-21
02087462 Rev. V
Understanding System Behavior with Custom Panels
When you select a custom panel, the system responds as follows:
•
The system disables any functions that you cannot use with custom panels.
•
If you select a custom panel that contains the tHb parameter and then select
Microsample, the tHb parameter is removed from the custom panel while
performing that analysis.
•
If you select a custom panel that does not contain the
p
O
2
and
p
CO
2
parameters and then select Microsample, the system does not perform the gas
measurement phase of the microsample measurement.
The Analysis screen does not display custom panels under some conditions:
•
you did not define custom panels in Setup
•
if you select a sample type other than a patient sample type, such as QC,
AQC, or High G/L
•
if you select pH only, pH Glu Lac only, or tHb only modes
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...