POWERMOBIL
SPR2-240.841.01
Page 6 of 6
Siemens AG
Rev. 01
05.03
CS PS 24
Medical Solutions
4 - 6
Checks and Adjustments
Braking forces / braking moments
4
•
Check the braking forces/braking moments according to the service instructions
(chapter 7, Braking forces / braking moments).
POWERPHOS settings
4
•
Set the POWERPHOS and collimator according to the system service instructions
(chapter 6, "POWERPHOS").
•
Afterwards, refit the POWERPHOS cover.
Covers
4
•
Refit all covers. Make sure the protective ground wires make good contact.
Image quality
4
Perform the IQ quick test.
Ground Wire Test
4
•
Perform the protective ground wire test according to ARTD-002.731.17.
➪
The protective round wire resistance must not exceed 0.2 ohms.