Checks and Adjustments
4
Setting the center position of the cable module to the
orbital center position of the C-arm
4
•
The lock of the cable module is released, the screw (1/Fig. 2) is in position (2/Fig. 2).
The C-arm can be moved orbitally.
•
Turn the C-arm orbitally so that the markings (1/Fig. 1) are centered on the threaded
holes (2/Fig. 1). The cable drums are in center position.
•
Lock the cable module in this position. For this purpose, turn the screw (1/Fig. 2) into po-
sition (1/Fig. 2).
•
Loosen the screw (1/Fig. 3), but do not screw it out. Press the screw (1/Fig. 3) in direc-
tion of the arrow (Fig. 3). With this the cable drums are uncoupled from the toothed belt of
the C-arm. The C-arm can be moved slightly orbitally.
•
Move the C-arm orbitally so that both markings (1/Fig. 4) on the toothed belt are posi-
tioned between the two rollers (2/Fig. 4). The C-arm is now in orbital 0° position.