Maintenance, Operating Value / Functional Inspection
39
Page 39 of 44
Medical Solutions
4.3 Functional
inspection
PMF
Video printer
•
Check the function of the video recorder.
•
Check the remote control (Print button).
PMF
Video recorder
•
Check the function of the video recorder.
•
Check the function of the remote control (REC-Pause-Start / REC-Pause-Stop).
PMF
DICOM bridge
•
Check the function of the DICOM bridge.
PMF
DICOM Connect
•
Check the function of the DICOM Connect network connection.
PMF
Codonics printer manual control panel
•
Check the function of the printer.
•
Check the function of the manual control panel.
•
Inspect the cable of the manual control panel for damage and breaks.
PMF
Laser camera connection
•
Check the functioning of the laser camera connection.
•
Check the remote control (Print button).
PMF
Image rotation
Check the image rotation (camera rotation).
X
•
To do this, check the 0° position and the entire rotational range (
±
220°) under fluoros-
copy.
•
Check the preview function of the image rotation.
PMF
Slot diaphragm
•
Select I.I. full format.
X
•
Open the slot diaphragm and briefly release fluoroscopy.
The slot diaphragm plates are no longer visible on the screen.
•
Select I.I. zoom format.
X
•
Release fluoroscopy briefly.
The slot diaphragm plates are no longer visible on the edges of the image.
X
•
Close the slot diaphragm.
Rotate the slot diaphragm
±
360° under fluoroscopy. The slot diaphragm can be rotated
360°.
PMF
Monitor display of the iris collimator aperture
•
Select I.I. full format.
•
Close the iris collimator (X-iris) completely.
•
Mark the diameter of the collimated iris collimator aperture on the monitor.