Medical Solutions
•
Release the orbital brake. The LED in the button goes on.
•
The servo support of the orbital movement must not operate. The C-arm may move in
the orbital direction only when increased force is exerted.
•
Press the Emergency Stop switch. The lifting column can be moved again.
•
Release the orbital brake.
•
The servo support function of the orbital movement is functional once again.. The
C-arm may move in the orbital direction with minimal force.
3.1.1
I.I. laser light localizer
SIM
Mechanical system
•
Inspect the I.I. laser light localizer for mechanical damage.
•
Mount the I.I. laser light localizer on the I.I. and ensure proper locking and seating.
When doing this, pay special attention to the tension band and its closure.
•
Check the I.I. ring for damage.
SIM
Function
•
Perform maintenance on the I.I. laser light localizer according to Installation and Set-
ting Instructions RXR2-130.033.03...
3.1.2 DICOM
bridge
SIM
Mounting
•
Check the mounting of the DICOM bridge in the monitor trolley.
3.1.3 Codonics
printer
SIM
Mounting
•
Inspect the screws used to fasten the guide rail to the mounting bracket of the
Framegrabber.
•
Ensure that the printer is properly fastened to the guide rail. Make sure there are no
missing attachment screws.
•
Ensure that the Framegrabber is properly fastened to the mounting bracket. Make sure
there are no missing attachment screws.
SIM
Guide rails
•
Inspect the guide rails for proper function and smooth operation.
•
The guide rails must not be damaged.
•
The guide rails must lock properly into place when fully pushed into position.
•
If necessary, use a cleaning brush to remove dirt from the rails.
3.1.4 Monitors/displays
SIM
Monitor/displays