7ML19985GG03
Pointek CLS500 – INSTRUCTION MANUAL
Page 79
m
mmmm
Appendix A
1. Set the rotary switch to 8.
2. Press and hold the RED (+) button for more than one second to select FSH,
or: Press and hold the BLUE (–) button to select FSL.
(In Free Programming Mode only, press and hold both buttons to disable this
function).
Miscellaneous
Output Signal Processing Test
Displays the Fault/Failure information. If operation is normal, two test displays alternate,
which light up all the LCD segments in a cycle. If there is a fault or failure, an error code is
displayed. See the detailed list of error codes and their meanings on page 60.
Set the rotary switch to 1.
To change the state of the output signals, press and hold both buttons simultaneously: the
digital mode outputs (the solid-state switch and the loop-current control in digital mode)
change to their opposite state. Thus 4mA becomes 20mA and contact open becomes
contact closed. This feature allows you to verify that the output signals are properly
processed further on in the PLC/DCS system: if the normal state is non-alarm, changing
the state should generate an alarm.
The outputs stay in the opposite state as long as both buttons are pressed.
Menu
Item
Rotary
Switch
Position
Left
Arrow
Mode
Description Action
Values
18
8
On
Digital
Signalling
Mode
*
FailSafe High
Press and hold
RED (+) button
Display
shows
FSH
FailSafe Low
Press and hold
BLUE (–) button
Display
shows
FSL
Menu
Item
Rotary
Switch
Position
Left Arrow
Description
11
1
On
Output Signal Processing Test