7ML19985GG03
Pointek CLS500 – INSTRUCTION MANUAL
Page 45
mm
mmm
User Int
e
rface
Adjusting the value
1.
Select a menu item.
2.
Press the RED (+) or BLUE (–) push button to adjust the value up or down:
acceptance is indicated when the left-most colon on the LCD appears. (You have to
press longer than the debounce time to have the action accepted; the debounce
time varies according to the menu item selected.)
Transmitter Variables
•
Transmitter Variable 0 (TV0) is the capacitance as measured by the device.
•
Transmitter Variable 1 (TV1) is a computed variable: the dynamic value is a
computed derivative from the range settings for TV0.
Notes:
•
The push buttons are preset with a delay called the debounce time.
•
Keeping one or both buttons pressed continuously will trigger an auto-repeat in
some menus.
Transmitter Variable User-defined Functions
Units
TV0
URV, LRV, Damping, USL and LSL
Fixed to pF
TV1
Can be user-defined