Page 32
Pointek CLS 500 – INSTRUCTION MANUAL
7ML19985GG03
mmmm
m
Int
e
rc
onnect
ion
Recommended Cable Entries (equivalents can be
used)
General Purpose Cable Entry: M20 (A5E03252531) and ½" NPT
(A5E03252530)
1.
Strip cable and expose braided shield.
2.
Feed cable through dome nut and clamping insert. Fold braided shield over clamping
insert. Make sure that braided shield overlaps the O-ring by 3/32" (2 mm) and covers
the entry 360 degrees.
3.
Push clamping insert into body and tighten dome nut. Assemble into housing.
Hazardous Location Cable Entry: M20 (A5E03252528) and ½" NPT
(A5E03252527)
1.
Strip the cable to suit equipment and expose armour/braid.
2.
Push the cable through the diaphragm shield/armour spigot (4). Pre-fitted cable
guide (4.1) can now be discarded. The diaphragm seal can be rolled back to ease
assembly as required. Spread armour/braid over the diaphragm seal/armour spigot
M20 x 1.5
6 mm (0.24")
24 mm
(0.94")
½" NPT
13 mm (0.51")
24 mm
(0.94")
½" NPT or
M20
15 mm (0.59")
64.4 mm
(2.54")
(1) Backnut
(2) Middle nut
(3) Reversible armour clamp
(4) Diaphragm seal/Armour spigot
(4.1) Cable guide
(5) Entry (with captive deluge seal)
(1)
(2)
(3)
(4)
(5)