7ML19985GG03
Pointek CLS500 – INSTRUCTION MANUAL
Page 5
mm
mmm
Specificat
io
ns
Technical Specifications: Pointek CLS500
Power
Supply voltage
• maximum
33 V DC
• minimum
12 V DC at 3.6 mA (9.5 V DC at 22 mA)
Loop current
3.6 to 22 mA / 22 to 3.6 mA (2-wire current loop)
Mounting
Location
indoor/outdoor
Altitude
2000 m max.
Ambient temperature
• standard
-40 to +85
o
C (-40 to +185
o
F)
• ATEX-Explosion Proof
-40 to +70
o
C (-40 to +158
o
F) for T6
-40 to +85
o
C (-40 to +185
o
F) for T5 to T1
Relative humidity
suitable for outdoor (Type 4X / NEMA 4X / IP65, IP68
enclosure)
Installation category
I
Pollution degree
4
Performance
Measurement range
Transmitter type
• MSP-2002-1
1 to 330 pF
Minimum span
1 pF
Measurement frequency
420 kHz @ Cx = 0 pF
Accuracy
deviation <0.1% of actual measurement value
Repeatability
0.1% actual measurement
Temperature stability
0.15 pF (0pF) or <0.25% (typically <0.1%) of actual
measurement value, whichever is greater over the full
temperature range of the transmitter